چکیده
|
In this research, a novel, rapid, and non-immune electrochemical method was used to detect cardiac troponin I
(cTnI) using a double recognition approach. Amine terminus cTnI aptamers immobilized on COOH-ZnO nanoparticles
(COOH-ZnONPs) modified GCE surface were applied to capture cTnI for imprinting recognition. The
COOH-ZnONPs were synthesized in a biological manner. Then, the methylene blue (MB) monomers were
electro-polymerized around the cTnI-aptamer complexes. Following the removal of cTnI, cavities were constructed
and converted to a new aptamer and molecular imprinted polymer (MIP) hybrid receptor (aptamer/
MIP/ZnONPs). FT-IR spectra, SEM images, XRD patterns, and electrochemical impedance spectroscopy (EIS)
were used to evaluate different steps. With respect to differential pulse voltammetry and the anodic peak of
polymethylene blue (PMB), cTnI was found to be in a linear range from 0.50-3.3×105 pM (1.25×10-5-8.25 ?g/
mL) with a detection limit of 1.04 pM (2.61×10-5?g/ml) and quantification limit of 1.16 pM (2.90×10-5?g/
mL). The selective determination of cTnI by the proposed aptamer-MIP/ZnONPs sensor was investigated in the
presence of cardiac troponin T (cTnT), C-reactive protein (CRP), myoglobin and human serum albumin (HSA).
The recovery of the electrode response to cTnI (5.00×103 pM) in a solution containing the proteins, as mentioned
above (5.00×103 pM) was about 91.2% (RSD=5.37%) with respect to cTnI alone.
|