چکیده
|
Heavy metal pollution of water samples changes the quality of water samples and these
contaminations have critical effects on human health. In this study, a modification of carbon
paste electrode was performed by MWCNTs and 3H-spiro[isobenzofuran-1, 6’-pyrrolo[2, 3-
d]pyrimidine]-2’,3,4’,5’-tetraones (3HSIPPT). Detection limit of 0.017, 0.011, and 0.003 mol
L-1and quantification limit of 0.031, 0.04, and 0.011mol L-1were achieved for Hg2+, Pb2+ and
Cd2+, respectively. The interference studies showed that the modified electrode is highly
selective to investigate the properties in the presence of metal ions. Therefore, the fabricated
sensors were effectively applied for simultaneous determination of analytes in several water
samples. Additionally, principal component analysis, k-mean clustering and hierarchical
clustering were applied for the classification studies using obtained metal content of target ions.
The DPV and flame atomic absorption spectrometry (FAAS) results were applied for clustering
of water samples. The water samples were classified into four groups using PCA. Comparison of
results of classification by different methods showed that all group members determined by PCA
and HC are in a good agreement with the predicted group of cases classified by k-mean
clustering. Therefore, the DPV results are applicable as an efficient technique to distinguish the
heavy metal pollution of water samples.
|